Background:
Atrio-esophageal fistula formation following radiofrequency ablation of left atrial tachyarrhythmias is a rare but devastating complication. Esophageal injuries are believed to be precursors of fistula formation and reported to occur in up to 47% of patients. This study investigates the incidence of esophageal lesions when real time esophageal temperature monitoring and temperature limitation is used.
Methods:
184 consecutive patients underwent open irrigated radiofrequency ablation of left atrial tachyarrhythmias. An esophageal temperature probe consisting of three independent thermocouples was used for temperature monitoring. A temperature limit of 40degreesC was defined to interrupt energy delivery. All patients underwent esophageal endoscopy the next day.
Results:
Endoscopy revealed ulcer formation in 3/184 patients (1.6%). No patient developed atrio-esophageal fistula. Patient and disease characteristics had no influence on ulcer formation. The temperature threshold of 40degreesC was reached in 157/184 patients. A temperature overshoot after cessation of energy delivery was observed frequently. The mean maximal temperature was 40.8degreesC. Using a multiple regression analysis creating a box lesion that implies superior- and inferior lines at the posterior wall connecting the right and left encircling was an independent predictor of temperature. Six month follow-up showed an overall success rate of 78% documented as sinus rhythm in seven-day holter ECG.
Conclusion:
Limitation of esophageal temperature to 40degreesC is associated with the lowest incidence of esophageal lesion formation published so far. This approach may contribute to increase the safety profile of radiofrequency ablation in the left atrium.
Read the original article at BMC Cardiovascular Disorders .